Divotech’s XR series of X-Ray sources are designed for OEM applications to power X-Ray tube up to 320kV at 1.6kW or 160kV at 800W or 125kV at 625W immersed in oil. Divotech’s power supplies are high frequency operated and compact in size.
Divotech’s X-Ray sources having an internal floating filament and a closed loop emission control circuit for precise regulation of emission current and control of Voltage & filament current.
If you need any help feel free to contact us. We are here to serve
Contact us
Input voltage | 230v AC +/- 10%, 50 Hz | |
High voltage supply | Output - 0 to +80 kV DC Output - 0 to -80 kV DC | |
Output voltage regulation | Line - 0.05% Load -0.1 % |
|
Output current | 0 to 5 mA DC | |
Output Voltage Stability | Within 0.1 % of set value | |
Output Voltage ripple | 0.2 % RMS on full load | |
Filament voltage and current | Small filament: 4V RMS - 4 Amps Large filament: 6V RMS - 4 Amps Variable filament voltage tapings are provided to match different X -Ray tubes. |
|
Remote Control settings | 0 to 10V DC for rated output voltage. 0 to 10V DC for rated X-Ray tube current. |
|
Remote Control readings | 0 to 10V DC for rated output voltage. 0 to 10V DC for rated X-Ray tube current. |
|
Control Connector (Rack to HV unit) | 1)7 Pin HITECH Connector. 2)24 Pin HITECH Connector. |
|
Protections | Over voltage Over current, short circuit& Arc |
|
Environmental | 0 to + 45 degrees at 10- 95% RH, non-condensing; Forced air cooling. | |
Dimensions for 160kV(800W) | Rack : 150mm (H)*380mm (W)*430mm (L) HV Unit: 250mm (H)* 320mm (w) *555mm (L) |
Remote control | 15 Pin "D" sub connector. | |
Input Power connector | Detachable 3 pin Power chord 250v Ac /16A. | |
15 Pin "D" sub connector connections | ||
Pin | Signal | Signal parameter |
1 | HV ON | |
2 | Filament selection | |
3 | V -set | |
4 | Ground | |
5 | I- set | |
6 | V- Monitor | |
7 | Ground | |
8 | I -Monitor | |
9 | NC | : Relay contact for over voltage |
10 | POLE | : Relay contact for over voltage |
11 | NO | : Relay contact for over voltage |
12 | NC | : Relay contact for over current |
13 | POLE | : Relay contact for over current |
14 | NO | : Relay contact for over current |
15 | DUMMY |